UV-Visible Spectrophotometric Method for Complexes Stoichiometry between Zn(II), Mg(II), Cd(II), Ca(II) and Cu(II) with Clonidine Hydrochloride

 

Serap Karaderi1*, Sena Bayraktar1, Caner Mazı1

1Depertmant of Analytical Chemistry, Faculty of Pharmacy, Marmara University-34668, Haydarpaşa, Istanbul, Turkey.

 

*Email: [email protected]


ABSTRACT

In the present study, an accurate, rapid, and simple, UV-Visible spectrophotometric method was developed for determining complexes stoichiometry of Zn(II)-clonidine hydrochloride, Mg(II)-clonidine hydrochloride, Cd(II)-clonidine hydrochloride, Ca(II)-clonidine hydrochloride, and Cu(II)-clonidine hydrochloride. The spectra of the complexes formed between clonidine hydrochloride with Zn (II), Mg (II), Cd (II), Ca (II), and Cu (II) cations were determined at 190-400 nm and the stoichiometry of the complexes formed between clonidine hydrochloride with the mentioned metal cations was determined at maximum wavelength using the mole ratio method. The stoichiometry of Cu2+-clonidine hydrochloride, Cd(II)-clonidine hydrochloride, Zn(II)-clonidine hydrochloride, and Mg(II)-clonidine hydrochloride were calculated as 1:1, but Ca(II)-clonidine hydrochloride, the complex was 1:2. The recovery was 100.8% for clonidine hydrochloride. The LOQ and LOD were 0.003 and 0.0028 µg/mL, respectively for clonidine hydrochloride. The interaction between clonidine hydrochloride with Cu2+, Ca2 + Cd2+, Mg2+, and Zn2+ were evaluated using the UV-vis spectrophotometric method, and reasonable results were obtained.

Key words:Clonidine hydrochloride, Cadmium(II), Copper(II), Zinc(II), Calcium(II), Magnesium(II)


INTRODUCTION

Clonidine hydrochloride Hypertension is a type of high blood pressure or cardiovascular disease. Antihypertensive agents can decrease high blood pressure [1]. Blood pressure consists of two valves, systolic and diastolic [2, 3]. Values greater than 140mmHg for SBP and 90mmHg for DBP are considered to be high blood pressure values, that is hypertension [4, 5]. Clonidine hydrochloride is N-(2,6-Dichlorophenyl)-4,5-dihydro-1H-imidazol-2-amine hydrochloride is a centrally acting α2 agonist drug that can inhibits stimulation of the sympathetic nervous system. In addition, Clonidine hydrochloride is an antihypertensive drugs used for menopausal flushing, migraine prophylaxis and high blood pressure.  This causes a decrease in heart rate and therefore decreased blood pressure [4, 6, 7].

The use of Clonidine hydrochloride is a type antihypertensive agent with a mechanism that differs from other common antihypertensive drugs. For example, Clonidine hydrochloride normalizes the blood pressure and clonidine hydrochloride is used for Tourette syndrome ADHD menopausal flushing, and migraine prophylaxis [8-11]. Clonidine hydrochloride lowers blood pressure by lowering certain blood chemicals. This causes the blood vessels to relax and the heart to slow down more slowly [12].

A low-dose administration leads to more effective control of blood pressure in both standing and supine positions in severe or moderate hypertension [9]. Also, narcotics are utilized to improve withdrawal symptoms associated with the long-term use of nicotine, benzodiazepine and alcohol [13]. In the literature review, some analytical methods for clonidine hydrochloride have been identified, but no studies have been conducted on the determination of the stoichiometry of complexes between clonidine with Zn(II), Mg(II), Cd(II), Ca(II) and Cu(II).

MATERIALS AND METHODS

Chemicals and reagents

All reagents used were of analytical grade. The pure standard of clonidine hydrochloride was obtained from Sigma-Aldrich Co. (USA). Copper (II) nitrate trihydrate, zinc (II) nitrate hexahydrate, calcium (II) nitrate tetrahydrate, magnesium nitrate hexahydrate, cadmium (II) tetrahydrate, and methanol reagent were obtained from Merck (Germany).

Apparatus

This experimental study was performed on Shimadzu UV-2550 series spectrophotometer with a double-beam detector configuration. The absorption spectra of test and reference solutions were performed in a 1.0cm quartz cell (Germany) in a range of 190 to 400 nm.

Preparation of standard stock and titration solutions

Clonidine hydrochloride (5.0×10-3M): 0.0333 g of clonidine hydrochloride was taken into 50mL volumetric flask and dissolved in 25mL methanol and the volume was made up to 50mL distilled water.

Cu(NO3)2.3H2O  (5.0×10-3M): Copper (II) nitrate trihydrate was prepared by dissolving 0.0604g copper(II)nitrate trihydrate in 50 mL mixture of water-methanol (50:50 v/v).

Ca(NO3)2.4H2O  (5.0×10-3M): Calcium(II)nitrate tetrahydrate was prepared by dissolving 0.0590g calcium(II) nitrate tetrahydrate in 50 mL mixture of water-methanol (50:50 v/v).

Cd(NO3)2.4H2O  (5.0×10-3M): Cadmium(II)nitrate tetrahydrate was prepared by dissolving 0.0771g cadmium(II) nitrate tetrahydrate in 50 mL mixture of water-methanol (50:50 v/v).

Mg(NO3)2.6H2O (5.0×10-3M): Magnesium nitrate hexahydrate was prepared by dissolving 0.0641g magnesium nitrate hexahydrate in 50 mL mixture of methanol-water (50:50 v/v).

Zn(NO3)2.6H2O (5.0×10-3M): Zinc(II)nitrate hexahydrate was prepared by dissolving 0.0744g zinc(II)nitrate hexahydrate in 50 mL mixture of water-methanol (50:50 v/v).

 

Selection of analytical wavelength of clonidine hydrochloride

For the selection of analytical wavelengths, the working test solution of bupropion hydrochloride (2.0×10-3M) was scanned between 400-190 nm. The overlay spectrum of clonidine hydrochloride was recorded.

Selection of analytical wavelength of Cu2+-clonidine hydrochloride and Ca2+ clonidine hydrochloride

Several test solutions were made by mixing different volumes of 2.0×10-3M solution of each of the copper (II)nitrate trihydrate and 2.0×10-3M of clonidine hydrochloride. The absorbance was scanned between 400-190 nm to determine their maximum wavelength using UV-2550 double-beam UV-visible spectrophotometer. The wavelength of the maximum absorbance and spectra were recorded.

Selection of analytical wavelength of Cd2+-clonidine hydrochloride, Mg2+-clonidine hydrochloride, and Zn2+-clonidine hydrochloride

Similarly, the absorbance of working test solutions of Cd2+-clonidine hydrochloride, Mg2+ -clonidine hydrochloride, and Zn2+-clonidine hydrochloride complexes was scanned between 400 to 190 nm to determine their maximum wavelengths. The maximum wavelength and spectra for Cd2+-clonidine hydrochloride, Mg2+-clonidine hydrochloride, and Zn2+-clonidine hydrochloride complexes were recorded.

 

Determination of the complex stoichiometry

The stoichiometries of Cu2+-clonidine hydrochloride, Ca2+-clonidine hydrochloride, Cd2+-clonidine hydrochloride, Mg2+-clonidine hydrochloride, and Zn2+-clonidine hydrochloride complexes were determined by using the mole ratio method. To determine the Cu2+-clonidine hydrochloride complex, several solutions were prepared such that the ratio of copper to clonidine hydrochloride was between 0-2.5, and the absorbance of these solutions was measured on UV spectrophotometer at the determined maximum wavelength.

Then, a graphic was drawn using the values between the ligand-metal mole ratio with absorbance. Similar operations were carried out in the same manner in other metal ions.

Validation

Linearity and calibration curve

For linearity and calibration curve, the working solutions were prepared at 0.066, 0.133, 0.266, 0.399, 0.533, and 0.666 µg/mL concentrations, respectively. The absorption spectra of solutions were scanned on spectrophotometer in the UV range of 400 to 190 nm, and their absorbance was recorded. The calibration curve between absorbance and concentration was drawn.

Precision

Interday and intraday variations were determined by analyzing four solutions (0.120, 0.33, 0.467, and 0.599 µg/mL) of clonidine hydrochloride within the same day and 3 different days for a week. The statistical parameters such as relative error (%RE) and relative standard deviation (%RSD) were calculated using the results obtained.

Accuracy

AThe analytical method accuracy was evaluated by the percentage recovery experiment carried out at one level equal to 100. The known amount of standard clonidine hydrochloride solution was added to the samples; absorbance was recorded and reanalyzed by the proposed method. The recovery percentage was calculated using the following formula:

 

n (mg) = C (mole/liter)×(molecular weight) (50,0 mL/1 tablet) × (dilution factor) =…… mg/ 1 tablet

(1)

 

% Recovery = (Found amounts of clonidine hydrochloride) / (Known amounts of clonidine hydrochloride) × 100

(2)

LOD and LOQ

Limit of quantification (LOQ) and limit of detection (LOD) were calculated using the following formula:

 

LOD=C+3s and LOQ=C+10s (C: concentration, s: standard deviation)

(3)

RESULTS AND DISCUSSION

The proposed method was based on spectrophotometric and mole ratio method determination of clonidine hydrochloride with metal complexes in the UV area using the mixture of water-methanol (50:50 v/v) as a solvent.

The absorption spectra of clonidine hydrochloride, Cu2+-clonidine hydrochloride, Ca2+-clonidine hydrochloride, Cd2+-clonidine hydrochloride, Mg2+-clonidine hydrochloride, and Zn2+-clonidine hydrochloride were found in the range of 190 to 400 nm. The curves between the absorbance and the wavelength have been shown in Figures 1 and 2.

 

Figure 1. The spectra of clonidine hydrochloride (2,0´10-3 M) clonidine hydrochloride for lambert-beer was found to be compatible in the concentration range 0.066-0.666 µg/ml  (Figure 2) with correlation coefficient (r2) of 0.9992.

 

The linear equation of standard deviation, correlation coefficient and regression in linearity, intersection and slope were calculated and summarized in Table 1. Figure 3 is the linearity plot of clonidine hydrochloride and Figure 4 determines the composition of Ca2+-clonidine hydrochloride complex by mole ratio method.

Figure 2. Absorption spectra of five different concentrations Cd2+, Ca2+, Zn2+, Cu2+, Mg2+ (0,066-0,666 µg/mL) of clonidine hydrochloride

 

Figure 3. Linearity plot of clonidine hydrochloride

 

Table 1. The statistical results of clonidine hydrochloride calibration curve

Statistical parameters

Result

Wavelength

272 nm

236 nm

Linearity range

0.066-0.666 µg/mL

Regression linear equation

y=0.4566×-0.0405

r2: Correlation coefficient

0.9992

Sa: Standard deviation on an intercept

0.738

Sb: Standard deviation on the slope

0.234

 

The LOD and LOQ values for Clonidine hydrochloride were found to be 0.0028-0.0039 µg/mL, respectively. RSD% and RE% day-to-day measurement results were calculated and summarized in Table 2.

Table 2. Intraday and interday precision measurements results

Intraday measurement values

Interday measurement values

Taken µg/mL

Found µg/mL

x̄±s

µg/mL

RE%

RSD%

Found µg/mL

x̄±s

µg/mL

RE%

RSD%

0.12

0.150

0.167

0.171

0.16±0.011

-18.5

6.79

0.152

0.183

0.187

0.17±0.019

-12.56

10.91

0.333

0.276

0.278

0.285

0.28±0.005

-16.2

1.718

0.277

0.286

0.305

0.29±0.014

-13.21

4.84

0.466

0.404

0.408

0.411

0.41±0.004

-12.6

0.98

0.407

0.433

0.420

0.42±0.013

-97.21

3.09

0.599

0.509

0.514

0.519

0.51±0.005

-14.19

0.97

0.519

0.530

0.536

0.53±0.008

-11.85

1.63

 

Recovery for clonidine hydrochloride was found to be 100.8%. Recovery percentage results are shown in Table 3.

Table 3. Recovery study of clonidine hydrochloride

100%

Known amounts of clonidine hydrochloride (mg)

Found amounts of clonidine hydrochloride (mg)

Recovery

%

1

39.68

32.60

82

2

49.61

51.37

101.5

3

59.53

70.90

119.1

The mean recovery %

 

 

100.8

 

The stoichiometry of Cu2+-clonidine hydrochloride, Mg2+- clonidine hydrochloride, Cd2+- clonidine hydrochloride and Zn2+ clonidine hydrochloride complexes were calculated as 1:1, but Ca2+-clonidine hydrochloride complex was calculated as 1:2.

 

Figure 4. Determination of the composition of Ca2+-clonidine hydrochloride complex by mole ratio method

 

CONCLUSION

 

In summary, in our study, the stoichiometry of the complexes formed by clonidine hydrochloride with some transition metals (Cu(II), Cd(II), Zn(II), Ca(II) and Mg(II)) was determined. The stoichiometry of Cu(II)-clonidine hydrochloride, Cd(II)-clonidine hydrochloride, Zn(II)-clonidine hydrochloride and Mg(II)-clonidine hydrochloride were calculated as 1:1, but Ca(II)-clonidine hydrochloride, complex was 1:2.

For all the values, low relative errors (RE< ±15%), high correlation coefficient (r2=0,9992) and high percentage recovery 100.8% showed the high linear relationship between the predicted and actual concentrations. The % RSD was less than 2%, without any significant difference in values for interday and intraday precision, which indicated the reproducibility of the method with high precision.

The lack of statistically significant difference between the values of the validation parameter (precision, accuracy, linearity, LOD and LOQ) for clonidine hydrochloride proved the system’s suitability for the developed UV-Vis spectrophotometric method. Literature information about complexes occurring between clonidine hydrochloride and metal cations was obtained. As a result, the precision of the analytical method was further substantiating. The proposed spectrophotometric method was rapid simple and accurate. Literature information about complexes occurring between clonidine hydrochloride and metal cations was obtained. We think that this information will contribute to the literature.

ACKNOWLEDGMENTS : The authors are thankful to Marmara University Scientific Research Projects Commission (BAPKO) for its support the project.

CONFLICT OF INTEREST : None

FINANCIAL SUPPORT : This research work has been supported by research grants from Marmara University Scientific Research Project SAG-C-YLP-110718-0440 by Marmara University Scientific Research Projects Committee.

ETHICS STATEMENT : None

References

1.       Karakoç EV, Ömer Z, Can H. Approach to Hypertension in Primary Care. Smyrna Tıp Dergisi. 2012:62-7.

2.       Mazı C, Karaderi S, Arıöz F. Determination of UV-Vis Spectrophotometric Method of Metal Complexes Stoichiometry between Cu (II) and Zn (II) with Vilazodone Hydrochloride. Int J Pharm Res Allied Sci. 2018;7(3):146-52.

3.       Karaderi S, Bakla M, Mazı C. Determination of UV-Vis Spectrophotometric Method of Metal Complexes Stoichiometry between Cu (II), Ca (II), Cd (II), Mg (II) and Zn (II) with Bupropion Hydrochloride. Int J Pharm Res Allied Sci. 2019;8(2):1-9.

4.       Corciova A. Validated colorimetric assay of Clonidine Hydrochloride from pharmaceutical preparations. Iran J Pharm Res. 2016;15(1):149-56.

5.       Yılmaz DÇ, Pekin M. Determination of Stability Constants of Ibandronate Complexes with Ca (II), Mg (II) and Sr (II). Int J Pharm Res Allied Sci. 2019;8(2):111-7.

6.       Khushboom G, Shah K, Pankaj P. Development and Valdatıon of RP-HPLC Method for Sımultaneous Estımatıon of Clonidine Hydrochloride and Hydrochlorothiazide in Pharmaceutıcal formulation. Int Bull Drug Res. 2014;4(6):105-6.

7.       Danafar H. Method validation of clonidine hydrochloride in human plasma by LC-MS technique. Pharm Biomed Res. 2015;1(4):48-58.

8.       Dilek M, Nurol A. Primer Hipertansiyon Patogenezi. Türk Klin J Nephrol- Spec Top. 2017;10(1):1-7.

9.       Houston MC. Clonidine hydrochloride: Review of pharmacologic and clinical aspects. Prog Cardiovasc Dis. 1981;23(5):337-50.

10.    Karaderi SE, Mazi C, Yilmazbilek S. Determination of the Stability Constants Comlexes of Metformin Hydrochloride With Cu (II), Cd (II) and Zn (II). Int J Pharm Res Allied Sci. 2020;9(1):121-9.

11.    Classen AM, Wimbish GH, Kupiec TC. Stability of admixture containing morphine sulfate, bupivacaine hydrochloride, and clonidine hydrochloride in an implantable infusion system. J Pain Symptom Manage. 2004;28(6):603-11.

12.    Oktar S, İlhan S, Aksulu HE. Clonidine prevents development of hypertension in N (omega)-nitro-L-arginine-treated rats/Klonidin N (omega)-nitro-L-arjinin aracili hipertansiyon gelisimini onler. Anatol J Cardiol. 2008;8(2):104-11.

13.    Bergendahl HT. Clonidine in paediatric anaesthesia: Pharmacokinetic and pharmacodynamic aspects. Institutionen för kirurgisk vetenskap/Department of Surgical Science; 2002.